Nice to see.  Those rules are also in the current OSHA electrical code and the NEC code.  The violation I continually see is a power cord plugged into a power bar or vice versa.  There is a common misconception that a power bar is not an extensions cord.



Monona Rossol, M.S., M.F.A., Industrial Hygienist
President:  Arts, Crafts & Theater Safety, Inc.
Safety Officer: Local USA829, IATSE
181 Thompson St., #23
New York, NY 10012     212-777-0062
actsnyc**At_Symbol_Here**cs.com   www.artscraftstheatersafety.org

 


-----Original Message-----
From: TILAK CHANDRA <tilak.chandra**At_Symbol_Here**WISC.EDU>
To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Sent: Thu, May 19, 2016 10:56 am
Subject: Re: [DCHAS-L] Electrical options for inside fume hoods

Based on Underwriter Laboratories (UL) recommendations:
 
=B7         Power (PTap)s  are not intended to be permanently secured to building structures tables, work benches and any other structure.
=B7         PTs should not be used as a substitute for fixed wiring.
=B7         Power taps or extension cords should not be connected to each other. Doing this can overload the circuit creating a potential fire hazard.
=B7         Power taps are not designed for high power loads such as space heaters, refrigerators and microwave ovens, which can easily exceed the recommended ampere ratings on many power strips.
 
However they are always present inside the chemical fume hood to supply the power to various instruments.
 
Good luck.
 
Tilak
 
Tilak Chandra, Ph.D.
Chemical Safety Specialist
UW-Madison
Ph. 608-890-0255
       608-622-9761 (Cell)
 
"--our job in safety is to make the task happen, SAFELY; not to interfere with the work-" Neal Langerman
 
 
 
From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**med.cornell.edu] On Behalf Of George MCCALLION
Sent: Thursday, May 19, 2016 7:31 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Electrical options for inside fume hoods
 
DCHAS group members,

Does anyone know of any power strips that are rated for use within chemical fume hoods? Typically the "standard" power strips are not recommended for such use.

Thank you in advance for any feedback you can provide.

Sincerely,
George

*******************************************************
George D. McCallion
Sr. Process Development Scientist, Chemistry
Novasep, Inc.
23 Creek Circle
Boothwyn, PA 19061
Main: +1.610.494.0447
Direct: +1.484.361.6023
Mobile: +1.610.888.2436
Email: george.mccallion**At_Symbol_Here**novasep.com
URL: www.novasep.com

NOTICE : This communication may contain confidential and privileged information that is for the sole use of the intended recipient. Any viewing, copying or distribution of, or reliance on this message by unintended recipients is strictly prohibited. If you have received this message in error, please notify us immediately by replying to the message and deleting it from your computer.
*******************************************************

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.