Subject: Re: Fume hood fan replacement

Thoughts....
 
Assuming that former system (before requiring new fan) was acceptable, review new TAB documents showing same airflow and pressure conditions as e xisted in prior system.   This could be all that's nece ssary but it depends on factors not easily surmised in a simply DCHAS post.   You may want to perform your own representative field measurements/v erifications to ensure new system components have not affected speed of res ponse of VAV dampers, etc.  Nor is a representative full ASHRAE 1 10 test or more ever a bad exercise in my opinion.
 
Wishing you success,


Steve Crooks, MS, CIH, CSP
President
People, Property & Environmental Protection, Inc.
919.368.7976
http:/ /ppeppro.com
 
 
-------- Original Message --------
Subject: [DC HAS-L] Fume hood fan replacement
From: Marjorie Markopoulos <marjorie.markopoulos**At_Symbol_Here**WRIGHT.EDU >
Date: Tue, December 28, 2010 1:34 pm
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU

Hello--
Does anyone have any thoughts or policies for test requirements for a system
of VAV fume hoods after a fan replacement? Do you have a specific protocol
or is full ASHRAE 110 testing required or recommended? As installed? As
use d? Any thoughts or comments would be helpful.
Thanks,
Marjorie Markop oulos

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.