Robert,
I am not sure whether the following would help or hinder, as there are a number of agencies that claim some sort of requirement for fume hood performance and testing.
The most commonly cited agencies include The American Society of Heating, Refrigerating, and Air Conditioning Engineers (ASHRAE), who developed a ANSI/ASHRAE Standard 110 Method of Testing Performance of Laboratory Fume Hoods. Another standard is NSI/AIHA Standard Z9.5 Laboratory Ventilation.
As mentioned in earlier comments, NFPA 45, 2004 has a regulation that states in part:
8.8.7 Measuring Device for Hood Airflow.
8.8.7.1 A
measuring device for hood airflow shall be provided on each chemical
fume hood.
8.8.7.2 The measuring device for hood airflow shall be a
permanently
installed device and shall provide constant indication to the hood user
of
adequate or inadequate hood airflow.
NFPA sells their codes and standards from their web site at http://www.nfpa.org
Best regards,
Tim
From:
DCHAS-L
Discussion
List [mailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behalf Of Hilbrandt,
Robert
Sent: Thursday, July 16, 2009 11:04 AM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: [DCHAS-L] Cite Reference
Could you site the requirement for fume hood monitors. We are located in New York State (upstate) and are not aware of such a requirement.
Robert W. Hilbrandt Jr., RMS
Radiochemistry
Technical Director / QA Manager
Westchester
County Dept. Laboratories & Research
10
Dana Road, Valhalla, NY 10595
Tel:
914-231-1531 Fax: 914-231-4458
rwh2**At_Symbol_Here**westchestergov.com
Visit
us at http://blog.w
estchestergov.com/index.php?option=com_content&task=view&id=
492&Itemid=1009
Previous post | Top of Page | Next post