Robert,

 

I am not sure whether the following would help or hinder, as there are a number of agencies that claim some sort of requirement for fume hood performance and testing.

The most commonly cited agencies include The American Society of Heating, Refrigerating, and Air Conditioning Engineers (ASHRAE),  who developed a ANSI/ASHRAE Standard 110 Method of Testing Performance of Laboratory Fume Hoods. Another standard is NSI/AIHA Standard Z9.5 Laboratory Ventilation.

As mentioned in earlier comments, NFPA 45, 2004 has a regulation that states in part:

8.8.7 Measuring Device for Hood Airflow.

8.8.7.1 A measuring device for hood airflow shall be provided on each chemical fume hood.
8.8.7.2 The measuring device for hood airflow shall be a permanently installed device and shall provide constant indication to the hood user of adequate or inadequate hood airflow.

NFPA sells their codes and standards from their web site at http://www.nfpa.org

Best regards,

Tim

 

From: DCHAS-L Discussion List [mailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behalf Of Hilbrandt, Robert
Sent: Thursday, July 16, 2009 11:04 AM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: [DCHAS-L] Cite Reference

Could you site the requirement for fume hood monitors.  We are located in New York State (upstate) and are not aware of such a requirement.

Robert W. Hilbrandt Jr., RMS
Radiochemistry Technical Director / QA Manager
Westchester County Dept. Laboratories & Research
10 Dana Road, Valhalla, NY  10595
Tel: 914-231-1531  Fax: 914-231-4458
rwh2**At_Symbol_Here**westchestergov.com
Visit us at http://blog.w estchestergov.com/index.php?option=com_content&task=view&id= 492&Itemid=1009

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.