Ellen,

 

After the incident at the Georgia Sperm Bank, our PD asked us to assess any spaces where a possible deficiency may occur.   We basically made some assumptions involving complete dead air (we have had a number of campus-wide power outages), calculated the volume of the laboratory (or other space), and then the size of the Dewar.  If the calculations showed that a 120 or 160L Dewar could displace enough oxygen to result in a low oxygen atmosphere, we installed local monitoring.   This included the NMR facility, a few labs using Dewars to act as a nitrogen gas source (we do not have house nitrogen), various storage rooms, and, last but not least - the room that houses SMU Athletics' cryotherapy unit.

 

Frankly, as alluded to, much of this is overkill.  We have had storage rooms drop down to about 19.4% when Dewars are venting as part of their pressure relief systems.  We recently had an NMR quench (that is a whole other story and involves a compressed air cylinder transported too close to the magnet).   Even then, during quenching, the oxygen in the NMR room never dropped below about 20%.  

 

Regards,

 

Brandon S. Chance, MS, CCHO

Director of Environmental Health and Safety

Southern Methodist University 

 

 

From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**Princeton.EDU> on behalf of "Ellen M. Sweet" <ems325**At_Symbol_Here**CORNELL.EDU>
Reply-To: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**Princeton.EDU>
Date: Wednesday, June 2, 2021 at 12:56 PM
To: "DCHAS-L**At_Symbol_Here**Princeton.EDU" <DCHAS-L**At_Symbol_Here**Princeton.EDU>
Subject: [DCHAS-L] benchmarking O2 monitoring with cryogen use

 

[EXTERNAL SENDER]

Hi everyone,

ÔJust doing some benchmarking. My question is what does your institution's assessment look like in order to determine when oxygen deficiency monitoring is needed for a proposal to use a cryogen liquid in a lab?

 

I look forward to hearing responses.

Thanks, Ellen

 

Ellen Sweet

Laboratory Ventilation Specialist

Department of Environmental Health and Safety, Cornell University

American Chemical Society, Division of Chemical Health and Safety

315-730-8896

 

--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.