Hi,
Two situations:
1.
Temperature controllers that are equipped with a "Triac" are likely to create problems because if the "Triac" fail it will put the controller in full open position thus generating
problems.
2.
Do not connect a variable transformer to a temperature controllerÉ this will also create problems to your controller.
Hope this help
From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**Princeton.EDU>
On Behalf Of Samuella Sigmann
Sent: 2020-09-21 12:34 PM
To: DCHAS-L**At_Symbol_Here**Princeton.EDU
Subject: [DCHAS-L] Website Updates - Electrical Safety Alert - Followup
All -
Just as a followup to this information, we pulled 12 controllers from our Organic lab rooms last week. Seems to be a real issue with heating mantle controllers.
Sammye
--
******************************************************************************
Non-Tenure Track Faculty - Qualified to cook the meals and wash the dishes, but not eat at the table.
Samuella B. Sigmann, MS, NRCC-CHO
Immediate Past Chair,
ACS Division of Chemical Health & Safety, 2020
Senior Lecturer/Safety Committee Chair/Director of Stockroom
Chemistry
Appalachian State University
525 Rivers Street
Boone, NC 28608
Phone: 828 262 2755
Fax: 828 262 6558
Email: sigmannsb**At_Symbol_Here**appstate.edu
--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at
membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas
Previous post | Top of Page | Next post