--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchasSection 4.5.6; B6 of the Z358.1 requires delivery of tepid water. Alan mentioned that the definition of tepid water was in the Appendices, however I saw somewhere that the temperature range was moved to (or included in) the Definitions section. I don't have a copy of the standard so I will defer to him on that. I assume that the word "tepid" appears in the Standard itself either way, is that right, Alan?The 2003 and later versions of the International Plumbing Code (IPC) require the use of ANSI/ISEA Z358.1 compliant equipment when eyewash or showers are required (around caustic or corrosive materials).. See, for example, https://safetyequipment.org/wp-content/uploads/2015/05/EWS-Guide-Feb_2015.pdf So this incorporates the tepid requirement. And even if your state has not adopted the IPC, it is obviously a best practice issue when you think about potential liability.Most state and local governments have adopted the IPC but, based on personal experience. less than 5% of people who purchase eyewash or shower units from us opt for the the thermostatic mixing valve even though their local codes would apparently require it. TMV's can cost more than the eyewash unit itself so I understand why people balk, but I've also seen people try to keep stating under a safety shower in the wintertime with ice cold water pouring over them when they are in shock or are injured.For code adoption see https://www.iccsafe.org/wp-content/uploads/Code_Adoption_Maps.pdf and click on IPC.Rob Toreki======================================================Safety Emporium - Lab & Safety Supplies featuring brand namesyou know and trust. Visit us at http://www.SafetyEmporium.comesales**At_Symbol_Here**safetyemporium.com or toll-free: (866) 326-5412Fax: (856) 553-6154, PO Box 1003, Blackwood, NJ 08012
On Nov 8, 2019, at 6:14 PM, Michael Schaffer <MichaelS**At_Symbol_Here**PSYCHEMEDICS.COM> wrote:--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchasI have a question about eyewashes.Do you have to regulate the temperature on the eyewash? We are in California, and we are installing a new eye wash.I thought there were some recommendations/codes?Thank you so much.,,,,,Michael I. Schaffer, Ph.D.., F.A.B.F.T., NRCC-TCVP Laboratory OperationsPsychemedics Corporation5832 Uplander WayCulver City, CA 90230310-216-7776 (phone)310-216-6662 (Fax)(800) 522-7424 (toll free) Extension 140*** CONFIDENTIALITY NOTICE ***THIS ELECTRONIC MAIL TRANSMISSION AND THE DOCUMENTS ACCOMPANYING IT CONTAIN INFORMATION BEING SENT BY PSYCHEMEDICS CORPORATION AND MAY CONTAIN INFORMATION THAT IS PRIVILEGED, CONFIDENTIAL AND/OR EXEMPT FROM DISCLOSURE. The information is intended only for the use of the person(s) named above. If you are not the intended recipient, you are hereby notified that any disclosure, copying, distribution, or the taking of any action in reliance on the contents of this information is strictly prohibited. If you have received this transmission in error, please immediately notify us by telephone at (800) 522-7424 and delete/destroy all copies.. Your receipt of this message is not intended to waive confidentiality or any privilege. The above name and title are not intended to constitute any form of signature.From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU> On Behalf Of Alan Hall
Sent: Friday, November 8, 2019 1:15 PM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: Re: [DCHAS-L] How does your institution interpret ANSI 358.1-2014 and safety shower testing?Eric et al,I am a voting member of the ISEA (International Safety Equipment Association) EWS (Eyewash and Shower) Product Interest Group who periodically (every 5 years) revise the ANSI/ISEA Z38.1 Standard. We will be meeting on November 18, 2019 to go over the final potential changes to the Z358.1 2019 revision..The issue of activation and testing is frequently brought up by those who utilize the Standard.If you have results from the D-CHAS group, I'd like to receive them before the morning of November 18, 2019 so that I can bring them to the attention of the Revision Committee.AlanAlan H. Hall, M.D.Medical ToxicologistToxicology Consulting and Medical Translating ServicesAzle and Springtown, TXClinical Assistant ProfessorColorado School of Public HealthUniversity of Colorado-DenverDenver, COOn Fri, Nov 8, 2019 at 11:22 AM Eric Goff <Eric.Goff**At_Symbol_Here**lcps.org> wrote:Trying to get the best practices here.How does your institution interpret ANSI 358.1-2014 and safety shower testing?I know eyewash stations are 5 minutes per week.What is the frequency you use for testing showers and the duration?Thanks in advance.Best Regards,Eric W. GoffChemical & Lab Safety SpecialistThe Academies of Loudoun42075 Loudoun Academy DriveLeesburg, VA 20175571.252.1980Be aware of serendipitous events!The opportunity of a lifetime must be seized during the lifetime of the opportunity.--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas
Previous post | Top of Page | Next post