This set up sounds like the one I have in my lab. I worked with the manufacturer directly when specifying my unit because it wasn't an advertised option. Turns out they sell this particular configuration quite often, but don't advertise
it.
In mine, I have two separate stand-alone steel units (left and right). Although they are two separate units, the manufacturer made them look like a single integrated unit by ensuring the doors "meet" in the middle and by having a single
piece kick plate.
My recommendation if that you contact the manufacturer directly and speak with their design engineers. They may already have a retrofit kit, but don't advertise it. At the very least, they will be able to tell you what can and can't be
done based on the design of these particular units.
Thank you,
Yaritza Brinker
260.827.5402
From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>
On Behalf Of Melinda Box
Sent: Wednesday, October 9, 2019 4:26 PM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: [DCHAS-L] possible conversion of under-hood acid cabinet to flammables?
** External Email **
Hi, all,
I am searching on behalf of a lab that does a significant volume of fundamental organic synthesis. As a result, they need to store so many 4-L bottles of flammable solvent near their work space that they are using the acid storage cabinets
for flammables. (Under each hood there is one acid storage cabinet and one flammables storage cabinet.)
The acid storage cabinets are distinguishable by the vents on the doors and the PTFE lining of the walls and floor inside the cabinet. The flammables storage cabinets are designed with no vents (specifically on the cabinet doors) and ~1"
of material mounted in a metal sheeted case on each side of the cabinet, presumably something flame resistant or retardant. So I am wondering if others have experience with a conversion of this sort. (Replacing the whole cabinet would be excessively disruptive
and cost prohibitive.)
For example it is it within acceptable practice to just replace the doors to the cabinet so that flammables storage does not have vents? Or to be up to fire code, would one be required to also change the lining of the cabinet (also prohibitive)?
Thanks for any experience on this matter you might be willing to contribute,
--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at
membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas
The information
contained in this electronic transmission is private, confidential, the property of the sender, and intended for the use of the recipient(s),
only. If you are not the addressee, any disclosure, copying, distribution or use of this information for any purpose is strictly
prohibited. If you have received this information in error, please notify the sender,
YBrinker**At_Symbol_Here**fele.com, immediately by e-mail and then delete this message. Thank you.
[FE.EN.1]
Previous post | Top of Page | Next post