From: Benjamin G Owens <bowens**At_Symbol_Here**UNR.EDU>
Subject: Re: [DCHAS-L] 4 more Re: [DCHAS-L] Regulations for Snorkels
Date: Thu, 28 Feb 2019 01:37:18 +0000
Reply-To: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>
Message-ID: CY4PR01MB22145B2550DB1572DE480144CB750**At_Symbol_Here**CY4PR01MB2214.prod.exchangelabs.com
In-Reply-To <5110206075852182.WA.chris.jakoberarb.ca.gov**At_Symbol_Here**lists.princeton.edu>


As has been stated, there aren't specific test requirements for task exhaust (snorkels), but I would measure the face velocity when they are installed. I would then re-measure once or twice a year and compare to the initial face velocity as a means of tracking performance relative to the initial conditions. This will help alert you to maintenance/performance issues.

Regards,
Ben

----------------------------------
Ben Owens
EH&S Assistant Director, Laboratory Safety
Environmental Health and Safety Dept., MS 328
University of Nevada, Reno
775-327-5196 (office)
775-843-2113 (cell)

-----Original Message-----
From: ACS Division of Chemical Health and Safety On Behalf Of Chris Jakober
Sent: Wednesday, February 27, 2019 11:21 AM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: Re: [DCHAS-L] 4 more Re: [DCHAS-L] Regulations for Snorkels

Hi Sammye,

RE: "The project manager asked me today if there were any regulations for testing them because the lab supplier is only planning on testing the hoods."

If the snorkels are being used to control exposures, CalOSHA expects them to be performance certified annually. I'm not sure if that's an expectation in other states. The code reference is 8CCR5143(a)(5).

https://na01.safelinks.protection.outlook.com/?url=https%3A%2F%2Fwww.dir.ca.gov%2Ftitle8%2F5143.html&data=01%7C01%7Cbowens%40UNR.EDU%7Cfde7f665cf114f264c0608d69cedc2a2%7C523b4bfc0ebd4c03b2b96f6a17fd31d8%7C1&sdata=BNBKk0kj5Gbm0%2FrdFsAXYQ9bJCTxhu2Sml9%2FdisJG%2B0%3D&reserved=0

Hope this helps,
Chris

Chris Jakober, Ph.D.
Air Pollution Specialist
Monitoring and Laboratory Division
California Air Resources Board
1900 14th Street ‰?? Sacramento, CA 95811
chris.jakober**At_Symbol_Here**arb.ca.gov
O: 916-324-1279

---
For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas

---
For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org
Follow us on Twitter **At_Symbol_Here**acsdchas

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.