Hmmm... just curious, how do you do weekly flushing of eyewashes as per ANSI standard?  And the shower flushing and flow testing?


We did not do flush units on a new science building that came online 3 years ago.  This is all new to me.


Vivian Longacre 
Safety Training Specialist
Environmental Health & Safety
Cal Poly State University
San Luis Obispo, CA

Direct 805.756.6628
vlongacr**At_Symbol_Here**calpoly.edu

From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU> on behalf of Chance, Brandon <bchance**At_Symbol_Here**MAIL.SMU.EDU>
Sent: Monday, October 24, 2016 1:40:06 PM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: [DCHAS-L] Shower/Eyewash Station Audible and Visible Alarms
 
DCHASers,

During a recent lab renovation, the contractor presented to us wall mounted shower/eyewash stations that have both an audible alarm and visible strobe when activated.  The unit also has the ability to communicate back tot he BMS system upon activation.  While this all sounds fine and dandy, it has raised a few questions on my end.

1.  Is this a new standard being used for installations?  I definitely see some practical purposes for this in order to alert others that an emergency has occurred, but I have not come across this before. 
2.  This raises some design dilemmas on our end.  
A.  We are installing these as flush mounted systems, so the units strobe and audible alarms would be above the drop ceiling (or we would have to relocate the alarm indicators). 
B.  For obvious reasons, the contractor was originally spec-ing out installation on walls without electrical just for cost and ease of plumbing.  If alarms are necessary, power must be ran to the units, thus increasing scope and cost. 

In general, I was just curious as to what kind of new installs other universities were currently putting in place.  


Regards,

Brandon S. Chance, M.S., CCHO
Associate Director of Environmental Health and Safety
Office of Risk Management
Southern Methodist University 
PO Box 750231 | Dallas, TX  75275-0231
T) 214.768.2430 | M) 469-978-8664

"… our job in safety is to make the task happen, SAFELY; not to interfere with the work…” Neal Langerman


--- This e-mail is from DCHAS-L, the e-mail list of the ACS Division of Chemical Health and Safety. For more information about the list, contact the Divisional secretary at secretary**At_Symbol_Here**dchas.org

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.