Hi George!

 

This is an application-specific question and would depend on which generic UL, X-proof or lesser class would apply.  There is no UL–Fume Hood outlet Class for all fume hood interiors since such a wide variety of internal conditions exist inside hoods.  It is far easier to use standard outlets OUTSIDE the hood environment, since these are typically Class I Div. 1 (Expensive) or GFI type and operate in the room rather than the fume hood environment. Most modern hoods have standard plug wire routing paths out of the containment area to such exterior outlets.

 

I hope this helps,

 

Dr. Bob Haugen

Director of Product and Technology Development

Flow Sciences Inc.

2025 Mercantile Drive

Leland, NC 28451

 

Phone 910 332 4878

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**med.cornell.edu] On Behalf Of George MCCALLION
Sent: Thursday, May 19, 2016 8:31 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Electrical options for inside fume hoods

 

DCHAS group members,

Does anyone know of any power strips that are rated for use within chemical fume hoods? Typically the "standard" power strips are not recommended for such use.

Thank you in advance for any feedback you can provide.

Sincerely,
George

*******************************************************
George D. McCallion
Sr. Process Development Scientist, Chemistry
Novasep, Inc.
23 Creek Circle
Boothwyn, PA 19061
Main: +1.610.494.0447
Direct: +1.484.361.6023
Mobile: +1.610.888.2436
Email: george.mccallion**At_Symbol_Here**novasep.com
URL: https://antispam.roaringpenguin.com/canit/urlproxy.php?_q=aHR0cDovL3d3dy5ub3Zhc2VwLmNvbQ%3D%3D&_r=Zmxvd3NjaWVuY2VzLWNvbQ%3D%3D&_s=ZHJib2I%3D

NOTICE : This communication may contain confidential and privileged information that is for the sole use of the intended recipient. Any viewing, copying or distribution of, or reliance on this message by unintended recipients is strictly prohibited. If you have received this message in error, please notify us immediately by replying to the message and deleting it from your computer.
*******************************************************



Spam
Phish/Fraud
Not spam
Forget previous vote

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.