Janice,

 

I’ve used muffle furnaces for two purposes for a couple of decades:  Ashing organics out of samples and performing salt fusions for analysis.

 

Ashing is performed in a low oxygen environment and is done relatively slowly over the course of 12-36 hours.  The door must be kept shut during the initial phases of the ashing to prevent deflagration.  A rapid introduction of oxygen at temperatures exceeding about 200-250 C can cause a deflagration.  Under no circumstances should the door be opened during the ashing process to “check on it.”  No, No, No.  Put a sign on the door saying “DON’T OPEN THIS.”    Most modern furnaces (for at least 25+ years) have the ability to ramp the temperatures up and down. and have overtemp shut offs on them.  NB:  These furnaces are not “sealed’ and do allow a limited amount of air into them during the process.  It’s a “low oxygen” environment, not a “no oxygen” environment.  Operation for ashing is usually an unattended operation after working hours; unless the lab director ignores the sign on the furnace door and opens it up.  Then it’s attended by the fire department among others.

 

Salt fusions are a completely attended operation as you have to do sample manipulations during the fusion process.

 

Harry

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**med.cornell.edu] On Behalf Of Dodge, Janice
Sent: Monday, April 25, 2016 11:48 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] high temp furnaces

 

I am trying to find out how other universities are handling the use of unattended high temperature furnaces that may need to run overnight or even up to several weeks.  In particular, I am interested in requirements for when these operations must be monitored continuously, the use of over-temperature shut down devices, and requirements related to flow-through gases. 

 

Thanks for any guidance or policies you may wish to share.

 

Janice

 

Janice Dodge

Laboratory Safety Officer

Florida State University

(850) 644-8916

 

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.