Why were the outlets on these not piped into the floor drain line or a drain line for a nearby sink?  This sounds like improper installation by the building contractor.

Digging up the floors to do it now would be a royal pain, but maybe you can connect into a lab sink drain line nearby.  I'd bring in your Physical Plant folks or a plumbing contractor to see if that's feasible.

Otherwise, the only possible way out I could think of would be to put some small basins under the outlet and rig up an inexpensive pump such as a air conditioning condensate pump, and run tubing along the wall to a sink.  A hack, and not a pretty one.

Are these Bradley or Guardian units?  If Guardian, send me some pics off list and I can get you a contact over there who might be able to offer ideas.

Rob

 ======================================================
Safety Emporium - Lab & Safety Supplies featuring brand names
you know and trust.  Visit us at http://www.SafetyEmporium.com
esales**At_Symbol_Here**safetyemporium.com  or toll-free: (866) 326-5412
Fax: (856) 553-6154, PO Box 1003, Blackwood, NJ 08012




On Aug 5, 2015, at 8:24 AM, "Stuart, Ralph" <Ralph.Stuart**At_Symbol_Here**KEENE.EDU> wrote:

Another question from this week's lab inspections:

We have a 2004 Science building with eyewashes built into the wall (similar to the design found at
https://urldefense.proofpoint.com/v2/url?u=http-3A__www.eyewashdirect.com_Bradley-2DS19294HB-2DBarrier-2DFree-2DEyewash-2Dw-2DTailpiece-2Dp_s19294hb..htm&d=BQIGaQ&c=lb62iw4YL4RFalcE2hQUQealT9-RXrryqt9KZX2qu2s&r=meWM1Buqv4IQ27AlK1OJRjcQl09S1Zta6YXKalY_Io0&m=EVo3tbkOvis_mZs4Ub41BYWHRcWLULcsLsexjSt5GMA&s=xM9PVi18EaMDhCtVyzxv7LI9sjPW1BE7hK0Sp0SB0fA&e=  ). When the eyewash is activated, the water flows back into the drain in the wall and then expelled onto the floor under the eyewash in an ever-growing puddle. My immediate concern is that the water comes out of a pipe which is between 6 and 10 inches from the floor. (There is a floor drain in the area of the pipe which serves the safety shower as well.) The imedidate problem is that location of the drain pipe is low enough that collecting water from it when flushing the eyewash is difficult; we currently use a secondary containment bin tipped under the pipe in such a way that we're able to collect about 1/3 of the capacity of th!
e bin.

I wonder if anyone on the list has developed a modification to this design that is reasonably priced (we're talking about 20 units with this condition) that would allow for collection of the flush water in a way that is convenient for the person conducting the flushing? This design also presents challenges for the emergency eyewash use case, but that's not my primary motivation in considering modifications; I'm more concerned with enabling more frequent flushing of the eyewashes.

Thanks for any experience with this question.

- Ralph

Ralph Stuart, CIH, CCHO
Chemical Hygiene Officer
Keene State College

ralph.stuart**At_Symbol_Here**keene.edu






Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.