From: "Czerwinski, Kevin" <kczerwin**At_Symbol_Here**UWSP.EDU>
Subject: Re: [DCHAS-L] Fume Hoods
Date: Sat, 1 Aug 2015 12:33:11 +0000
Reply-To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Message-ID: 47435E55-7647-4D3D-86FD-3203E0DFCFD0**At_Symbol_Here**uwsp.edu
In-Reply-To


Also depends on jurisdiction. State facility in a non-OSHA state? For example, state law in Wisconsin is 100 fpm.


Dr. Kevin M. Czerwinski, Ph.D.
Professor
Department of Chemistry

University Chemical Hygiene Officer
Environmental Health and Safety

B150 Science Building
University of Wisconsin-Stevens Point
2001 Fourth Avenue
Stevens Point, WI 54481

715-346-4154 (Office)
715-340-2216 (Mobile)

> On Aug 1, 2015, at 7:20 AM, Ellen M. Sweet wrote:
>
> Kevin,
> The 80-120 fpm range, from the ANSI Z9.5, is based on performance (containment). The standard also states that some hoods will perform at lower face velocities. Your organization needs to decide what face velocity you want your hoods at, unless you're in California.
> 100 fpm is typical.
>
> Ellen
>
> -----Original Message-----
> From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Kevin Burns
> Sent: Friday, July 31, 2015 10:17 AM
> To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
> Subject: [DCHAS-L] Fume Hoods
>
> I am currently finding myself in trying to deal with a fume hood problem. We had a mechanical problem that needed to be repaired causing The system to be shut down. Now that the system can be turned back on we need to have an air balancer come in and rebalance the system. After that my certification company would come in to certify that the hoods are maintaining proper CFMs for employees to work in them within the laboratories. The air balancer is in disagreement with the certification vendor. The air balancer says we have to have a policy that our face velocities for each hood must be 100cfm, where the certifiers are indicating that it's between 80 and 120 which that was what my understanding was.
>
> I know this falls under ASHRAE standards, ANSI standards and OSHA standards, can you point me in the right direction of where to get the exact information?
>
> Thank you, any information would greatly be appreciated.
>
> KB
>
> Kevin Burns
> Sent from my iPhone

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.