From: Louis DiBerardinis <loudib**At_Symbol_Here**MIT.EDU>
Subject: Re: [DCHAS-L] Particulate protection from a milling machine
Date: Thu, 5 Mar 2015 12:30:26 +0000
Reply-To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Message-ID: B0293D9281F2FF48A8D680CCDCE21166010DC6D610**At_Symbol_Here**OC11EXPO33.exchange.mit.edu
In-Reply-To <852F486739ECB34484E384E1D009D97D2B41445C**At_Symbol_Here**UBOX2.unr.edu>


Ben is correct in that “fume “ hoods is an incorrect term that many have been trying to change for years. It is designed to contain vapors and gases not particulates or fumes. The transport velocity needed would be 3-4000 fpm in the duct and  most systems are designed for a tduct velocity of 1500 fpm. So particles will settle out and clog the system.

 

The ACGIH vent manual describes these conditions. Another valuable reference is ANSI standard Z9.5-2012“ laboratory ventilation”. Several colleagues and I also have a book, Guidelines for Lab design, 4th edition from Wiley ( or on amazon much cheaper)that focuses more on the health and safety aspects of labs.

 

Lou DiBerardinis

Director, EHS office

MIT

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Benjamin G Owens
Sent: Wednesday, March 04, 2015 4:11 PM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: Re: [DCHAS-L] Particulate protection from a milling machine

 

Suzy,

 

When designing a ventilation system to capture and transport toxic particulate such as lead particulate, a particulate filter (perhaps a HEPA) will generally be located as close as possible to the capture point to minimize contamination of the ducting and HVAC system.  Proper transport velocity Is also required to prevent settling out of the particulate.  One commonly used ventilation design guide is the ACGIH publication Industrial Ventilation: A Manual of Recommended Practice for Design (http://www.acgih.org/store/ProductDetail.cfm?id=2235). 

 

Regards,

Ben

--------------------------------

Ben Owens

Assistant Director, Laboratory Safety

Environmental Health and Safety Dept., MS 328

University of Nevada, Reno 89557

Office Phone: 775-327-5196

Cell Phone: 775-843-2113

Fax: 775-784-4553

 

 

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Suzy Arnette
Sent: Wednesday, March 04, 2015 11:58 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Particulate protection from a milling machine

 

Good afternoon,

 

We have a lab that uses their milling machine in the chemical fume hood to protect the researchers from lead particulates in the material they are working on.  We are now getting push back from our HVAC folks about this process because they feel the particulates are clogging our heat capture coils.  

 

Do you any of you have similar processes?  Have you received push back from HVAC?

 

Thank you,

Suzy

 

--

Suzy Arnette, PhD
Lab Safety and Design Manager
Radiation Safety Officer
Boise State University
Environmental Health, Safety, and Sustainability
Phone: 208.426.3906  
MS-1826 
http://operations.boisestate.edu/EHS 


Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.