From: Eric Clark <erclark**At_Symbol_Here**ph.lacounty.gov>
Subject: Re: [DCHAS-L] House vacuum systems - traps
Date: Mon, 20 Oct 2014 21:39:27 +0000
Reply-To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Message-ID: 17A66C0B22391144A0BEE1CA471703EA77BB0873**At_Symbol_Here**ITSSDOWEXMB11.HOSTED.LAC.COM
In-Reply-To


For a stand-alone vacuum system with a single vacuum pump, like for a distillation apparatus or schlenk line it would be pretty much essential to put a cold trap in the vacuum line so you don’t risk messing up a $5000 vacuum pump. 

 

A house vacuum system generally has a built-in trap prior to the vacuum ballast tank, as is the case at this facility.  If a procedure were to involve something that would put the house system at risk for damage like if something bad were to inadvertently get sucked up, then some kind of precautionary trap or shut-off would be put in place and it would be written up like that in the specific SOP.  But we don’t run anything here that would risk damaging the house vacuum pump system so we don’t have to take that precaution. 

 

BTW – is anyone still using those horrible water-wasting noisy messy sink faucet vacuum aspirators (requires 6 – 12 liters/minute OMG!)?  I’m happy to report we do not. 

Eric

 

Eric Clark, MS, CHMM, CCHO

Safety Officer, Public Health Scientist III

Los Angeles County Public Health Laboratory

 

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Brandon S. Chance
Sent: Monday, October 20, 2014 12:07 PM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: Re: [DCHAS-L] House vacuum systems - traps

 

I thought I would try putting this message out there one more time and hope for the best. 

 

Brandon

 

 

 

Dear DCHAS,

 

Are any of you running building traps on your house vacuum systems?  I know standard procedures require lab users to use traps when tapping into the system, but I am curious if any of you have went a step further and installed traps within your mechanical spaces that house the vacuum pumps.  If so, could you please provide me additional information on the traps you are using.

 

Regards,

 

 

Brandon S. Chance, M.S., CCHO

Program Manager, Chemical Safety

Environmental Health and Safety

Princeton University

262 Alexander Street

Princeton, NJ 08540

609-258-7882 (office)

609-955-1289 (mobile)

609-258-1804 (fax)

 

"… our job in safety is to make the task happen, SAFELY; not to interfere with the work…” Neal Langerman

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.