Thank you. This is a great example of crowdsourcing for information. Importantly, the storage under an inert gas is not primarily for safety purposes.
Assuming this is the best answer to Ralph’s question, could someone please add this information to Wikipedia or ChemSpider? (It may already be on ChemSpider somewhere.)
Pete Reinhardt, Yale EHS
From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU]
On Behalf Of Slawomir Janicki
Sent: Saturday, April 12, 2014 12:08 PM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: Re: [DCHAS-L] Dichloromethane storage?
Dichloromethane can react with air moisture and oxygen and produce some nasty impurities like phosgene, hydrogen chloride, chloroform and 1,1,2,2-tetrachloroethane.
DCM is more stable than chloroform, but not stable enough to leave it without inerting for storage.
If you google “dichloromethane decomposition” you will get many references with more details.
Slawomir Janicki, Ph.D.
From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU]
On Behalf Of Monona Rossol
Sent: Saturday, April 12, 2014 7:03 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: Re: [DCHAS-L] Dichloromethane storage?
Whether the gases above the liquid are air or inert gases, the evaporation rate should be the same at the same temperatures and pressures. So that's not the reason.
Monona Rossol, M.S., M.F.A., Industrial Hygienist
President: Arts, Crafts & Theater Safety, Inc.
Safety Officer: Local USA829, IATSE
181 Thompson St., #23
New York, NY 10012 212-777-0062
-----Original Message-----
From: Jim O'Connor <oconnor.jim77**At_Symbol_Here**GMAIL.COM>
To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Sent: Sat, Apr 12, 2014 6:06 am
Subject: Re: [DCHAS-L] Dichloromethane storage?
Sigma Aldrich has a SDS that says, "Heat sensitive. Store under inert gas" I
don't agree with that statement but thought you should know about the ref. My
experience says the inert gas recommendation is to prevent it from evaporating
while storing. A liquid seal cap should negate the need for the inert gas too I
think.
Sent from my iPhone
> On Apr 11, 2014, at 12:39 PM, "Ralph B. Stuart" <ralph.stuart**At_Symbol_Here**CORNELL.EDU>
wrote:
>
> A lab tech noticed a statement on a Dichloromethane bottle that said "store
under an inert gas". He would like to know if this is a safety recommendation or
designed to maintain the quality of the chemical by avoiding exposure to
environmental gases. From what I can find, it appears to be the latter, but I
thought I would ask DCHAS-L to see if anyone knows of a safety reason for this
recommendation.
>
> Thanks for any help with this.
>
> - Ralph
>
>