From: Yung Morgan <pmorgan**At_Symbol_Here**EHS.UMASS.EDU>
Subject: Re: [DCHAS-L] Chemistry Fume Hood Experience
Date: April 29, 2013 4:57:21 PM EDT
Reply-To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Message-ID: <CDA2F208.19CA6%jsjsrj**At_Symbol_Here**comcast.net>


For the new high efficiency hoods,   require  an Ashrae  110  test  as manufactured(AM)   and   another one  as installed (AI) for  at least 10 or 20% of the hoods  in your facility. 

Air flow alarm  visible( with numbers)  and audible are a requirement..  When testing the hoods   we  have a required that the  hoods pass at flow rate: 80 FPM(70-90FPM) at 18 inches opening  and hood alarm set at 55FPM , sash response time at 60 sec or less.   Ask for a balancing report also to be sure the supply is proper for the space as a VAV hood  will not work without enough supplied air, hence the  constant flow alarm which can be a nuisance to the occupants.   Fume hoods sensors are not always useful  as research hoods  have to run overnight  at optimum flow rates, a constant battle between the energy group and safety people.  Room occupancy sensors are fine as long as the researchers know to not to work in the dark(with the light off)  unless they want their hoods air flow to go down.   

 

Again, just my two cents.

 

Yung Morgan, MsPH
Laboratory Safety
Industrial Hygiene Services
Environmental Health and Safety
117 Draper hall
UMASS,Amherst MA 01003
phone (413)  545-2682
Fax  (413) 545-2600
email : pmorgan**At_Symbol_Here**ehs.umass.edu

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Jim Johnson
Sent: Sunday, April 28, 2013 6:35 PM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Chemistry Fume Hood Experience

 

Good Afternoon,
I would like to start a discussion on current experience with chemistry fume hoods related to overall quality, best value, interior construction, coatings/materials of construction, installation issues, air flow alarms, order lead time, energy saving features plus anything else that comes to mind.
Thank you,

James S. Johnson Ph.D., CIH, QEP
JSJ and Associates
Pleasanton, CA 94588

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.