--20cf307f3b5e21bb4004b31abf38 The utility panel(s) are located inside the workplace along both sides of the hood and are used to access the plumbing for the water, air and gas outlets which are located next to the panel (s). On Thu, Dec 1, 2011 at 12:39 PM, Michael E. Richardson < mrichardson**At_Symbol_Here**flowsciences.com> wrote: Mary Ellen:**** What part of the fume hood are you referring to when you say =93utility panel=94? If you are talking about the back where the small openings are inside the work area, then yes you are absolutely right. I know this as the baffle (or with our enclosures the plenum). It is specifically designed for best air flow with this in place. **** ** ** V/R,**** Mike**** ------------------------------ *From:* DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] *On Behalf Of *Mary Ellen A Scott *Sent:* Thursday, December 01, 2011 10:30 AM *To:* DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU *Subject:* [DCHAS-L] Fume Hood**** ** ** Hello Everyone, When the utility panel on a chemical fume hood is missing is it safe to allow work to be preformed in the hood? I would say work (synthesis, toxic gas use and powder decanting) should not be preformed until the panel be replaced for the following reasons: 1. Eddyies in the hood are likely to be form which may cause loss of containment. 2. Chemicals vapors are going into areas where they are not suppose go and may cause deterioration of the plumbing and joints 3. High Performance and Low flow hoods would be especially vulnerable to the loss of work space integrity. Would anyone have a policy on hood repairs like the above or comments? As always, thank you for your thoughts in advance. Mary Ellen **** -- Mary Ellen Scott, PhD. Safety Specialist II Case Western Reserve University EHS - Environmental Health and Safety Service Building 1st Floor Rm 113 2220 Circle Dr. Cleveland, OH 44106-7227 216-368-6077 216-368-2236 (Fax) maryellen.scott**At_Symbol_Here**case.edu *=93There is no science without fancy and no art without fact=94 =96 Vladimir Nabokov* (1899-1977) --20cf307f3b5e21bb4004b31abf38 The utility panel(s) are located inside the workplace along both sides of the hood and are used to access the plumbing for the water, air and gas outlets which are located next to the panel (s).On Thu, Dec 1, 2011 at 12:39 PM, Michael E. Richardson <mrichardson**At_Symbol_Here**flowsciences.com> wrote:Mary Ellen:
What part of the fume hood are you referring to when you say =93utility panel=94? If you are talking about the back where the small openings are inside the work area, then yes you are absolutely right. I know this as the baffle (or with our enclosures the plenum). It is specifically designed for best air flow with this in place.
V/R,
Mike
From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Mary Ellen A Scott
Sent: Thursday, December 01,2011 10:30 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Fume Hood
Hello Everyone,
When the utility panel on a chemical fume hood is missing is it safe to allow work to be preformed in the hood?
I would say work (synthesis, toxic gas use and powder decanting) should notbe preformed until the panel be replaced for the following reasons:
1. Eddyies in the hood are likely to be form which may cause loss of containment.
2. Chemicals vapors are going into areas where they are not suppose go and may cause deterioration of the plumbing and joints
3. High Performance and Low flow hoods would be especially vulnerable to the loss of work space integrity.
Would anyone have a policy on hood repairs like the above or comments?
As always, thank you for your thoughts in advance.
Mary Ellen
--
Mary Ellen Scott, PhD.
Safety Specialist II
Case Western Reserve University
EHS - Environmental Health and Safety
Service Building 1st Floor Rm 113
2220 Circle Dr.
Cleveland, OH 44106-7227
216-368-6077
216-368-2236 (Fax)
maryellen.scott**At_Symbol_Here**case.edu
=93There is no science without fancy and no art without fact=94 =96 Vladimir Nabokov (1899-1977)
--20cf307f3b5e21bb4004b31abf38--
Previous post | Top of Page | Next post