Date: Thu, 1 Dec 2011 19:29:07 +0000
Reply-To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Sender: DCHAS-L Discussion List <dchas-l**At_Symbol_Here**med.cornell.edu>
From: "Michael E. Richardson" <mrichardson**At_Symbol_Here**FLOWSCIENCES.COM>
Subject: Re: Fume Hood
X-To: DCHAS-L
In-Reply-To: <CAKoY-_z1Ao+SOkbzVf=KCtxVFBOz7vN1dUh=BM-T6-aRib8PEQ**At_Symbol_Here**mail.gmail.com>

Mary Ellen:

   Most manufacturers of fume hoods and safety enclosures use different terms for the parts of a hood. Cornell has a .pdf that lists the parts and their function ( http://sp.ehs.cornell.edu/lab-research-safety/laboratory-safety-manual/Documents/SafeHoodsUseGuide.pdf ). If you mean the baffles instead of utility panel you are very much correct. Proper containment and airflow can onlybe gotten with all of the parts that affect the airflow in place.

 

Best regards,

 

Michael E. Richardson

Lab Manager

Flow Sciences, Inc.
mrichardson**At_Symbol_Here**flowsciences.com
corporate office 800-849-3429 Ext. 4849
corporate fax    910-763-1220
www.flowsciences.com
2025 Mercantile Drive
Leland, NC 28451

size=2 width="100%" align=center tabindex=-1>

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU] On Behalf Of Mary Ellen A Scott
Sent: Thursday, December 01,2011 10:30 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] Fume Hood

 

Hello Everyone,

When the utility panel on a chemical fume hood is missing is it safe to allow work to be preformed in the hood?

I would say work (synthesis, toxic gas use and powder decanting) should notbe preformed until the panel be replaced for the following reasons:

1. Eddyies in the hood are likely to be form which may cause loss of containment.
2. Chemicals vapors are going into areas where they are not suppose go and may cause deterioration of the plumbing and joints
3. High Performance and Low flow hoods would be especially vulnerable to the loss of work space integrity.

Would anyone have a policy on hood repairs like the above or comments? 
As always, thank you for your thoughts in advance.

Mary Ellen

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.