Hello Everyone,

When the utility panel on a chemical fume hood is missing is it safe to allow work to be preformed in the hood?

I would say work (synthesis, toxic gas use and powder decanting) should not be preformed until the panel be replaced for the following reasons:

1. Eddyies in the hood are likely to be form which may cause loss of containment.
2. Chemicals vapors are going into areas where they are not suppose go and may cause deterioration of the plumbing and joints
3. HighPerformance and Low flow hoods would be especially vulnerable to the loss of work space integrity.

Would anyone have a policy on hood repairs like the above or comments?
As always, thank you for your thoughts in advance.

Mary Ellen



Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.