Hi Angelo:
Here are a few excerptsfrom
NFPA that should help:
NFPA 70, Article 210 requires
GFCIs in these locations (I left out items (1)-(4) as they are not applicable
to labs):
(5) Sinks — where receptacles are installed within1.8
m
(6 ft) of the outside edge of the sink.
Exception No. 1 to (5): In industrial laboratories,
receptacles
used to supply equipment where removal of power
would introduce a greater hazard shall be permitted
to be
installed without GFCI protection.
Exception No. 2 to (5): For receptacles located in
patient
bed locations of general care or critical care areas
of
health care facilities other than those covered under
210.8(B)(1), GFCI protection shall not be required.
(6) Indoor wet locations
(7) Locker rooms with associated showering facilities
(8) Garages, service bays, and similar areas where electrical
diagnostic equipment, electrical hand tools, or portable
lighting equipment are to be used
NFPA 45 has the following
regarding fume hood electrical devices:
8.8.4* Electrical Devices.
8.8.4.1 In installations where services and controls
are within
the hood, additional electrical disconnects shall be located
within 15m(50 ft) of the hood and shall be accessible
and clearly
marked.
8.8.4.2 If electrical receptacles are located external
to the
hood, no additional electrical disconnect shall be required.
(See 5.6.1.)
Regards,
Dan
Dan Shiel, MBA, CSP, CHMM
Director, Risk & Quality
Management Services
Pfizer La Jolla/ GO R&D
West
10777 Science Center Drive
San Diego, CA 92121
“Money is the most envied,
but the least enjoyed.
Health is the most enjoyed, but the least envied.”
Charles Caleb Colton,1780-1832, British sportsman writer
From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**MED.CORNELL.EDU]
On Behalf Of Angelo T Chinni
Sent: Thursday, October 27, 2011 9:50 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] GFCI's and requirements for locating near safety
showers, and/or Sinks. . .
All,
(Please forgive me if this is a repeat question. . .)
I am sending a request to the group for assistance in locating any U.S.
regulations, design standards, codes with regard to providing Ground
Fault Circuit Interuptor (GFCI) protection (in Laboratories) for
electrical outlets located within the vicinity of eyewashes, and/or sinks.
Additionally, are laboratory fume hoods (with water and drains supplied)
provided with such protection (in addition to emergency power shut-off
switches).
I would appreciate hearing from this group's experience.
Any history, background or references would be much appreciated. Please
reply directly to chinniat**At_Symbol_Here**lilly.com
Thanks.
Angelo
Angelo T. Chinni, CIH, CSP
Mail Drop Code 4914
Phone: 317-277 8390
Lilly Corporate Center -LTC Fax:
317-651-6937
Indianapolis, IN 46285
Mobile: 317-730-0924
E-Mail: Chinniat**At_Symbol_Here**lilly.com
Previous post | Top of Page | Next post