Regina, I echo what’s been said about a “no entry
8221; zone around the magnet for anyone with medical devices that could be
impacted. We found that marking the floor was not enough to keep the
reminder fresh – you can get plastic stanchions and chains very inexp
ensively that help remind everyone about the barrier (also reminds folks to
remove watches, wallets and cell phones when approaching the magnet!).
The ot her tidbit would be to make sure you have adequate ventilation in the room – preferably install a permanent oxygen monitor and have it regularly calibrated and maintained. Depending on the configuration of your ro om and ventilation, refilling the liquid helium/nitrogen can create a poten tially hazardous atmosphere in the room, as can system leaks.
< /span>
Regards,
Misty L. Bogle
Corporate Health, Safety & Product Stewardship Manager
Vertellus Specialti es Inc.
201 N. Illinois Street, Suite 1800
Indianapolis, IN 46204
< /span>
From: DCHAS-L Discussion List [m
ailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behalf Of Regina Frasca
Sent:
b> Thursday, July 28, 2011 4:40 PM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: [Ninja][DCHAS-L] Lessons learned with NMR units.
span>
Colleagues,
Any good lessons le arned from the installation or operation of an NMR from a safety standpoint . We are just about to start installing a new unit and besides the ma nufacture specs and my CIH’s recommendations, I wanted to reach out there and see if anyone has any good tidbits.
Regina M. Frasca, NRCC-CHO
Cal State University San Marcos
Director of
Risk Management & Safety
Work: 760-750-4502
Fax:  
; 760-750-3208
rfrasca**At_Symbol_Here**csusm.edu
The FAS Division Transforms t
he Workplace…
Service, Laughter, Inviting, Excitement, Communicati
on, Innovation, Finest
RM&S Mi ssion: Contributing to a total learning environment by prov iding support, guidance, and leadership, promoting safety and preserv ing the human and physical resources of the campus community.
Previous post | Top of Page | Next post