Donna:
 
You're right, the hood sash can't be completely cl osed unless the block is moved out of the way.  It leaves a gap of a few in ches above the deflector shield.
 
In our case, there are other factors which, so far , have make this a workable fix:
 
1) we have a small staff (easy to train and monitor);
2) actual hood usage is minimal;
3) per our safety policy, we don't keep large volu mes of flammables/combustibles in open containers in the hoods; and< /DIV>
4) we have minimized the sources of ignition (open flame, spark sources).
 
As part of the shutdown procedure for overnight, a ll the blocks are removed and the sashes closed.
 
Rick Clark


From: DCHAS-L Discussion List [mailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behalf Of Majewski, Donna - West Lafayette, IN
Sent: Wednesday, November 11, 2009 5:48 PM
To : DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: Re: [DCHAS-L] Hood Sash Cables

Mr. Clark,

 

With the steel b lock in place, can you still fully close the sash?

 

Have a safe day,

 

Donna Majewski

Great Lakes Chem ical Corporation

(a Chemtura company)

EHS Regional Manager

1801 US Hwy 52 West

West Lafayette , IN  47906

Ph:  765-497-6254

Fax:  765-497-5995

 


From: DCHAS-L Discussion List [mailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behal f Of Clark, Richard C
Sent: Wednesday, November 11, 2009 2 :29 PM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: Re: [DCHAS-L] Hood Sash Cables

We had a cable f ail at the counterweight a few months ago.  That particular fume hood had onl y a single counterweight rather than one for each side of the sash.  Fortunately, no one was using the hood at the time.  The sash drop was not far and there was no glass breakage.  The hood is nearly twenty years old.  It would've caused injury; it took two people to raise the sash afterwards.

 

One quick soluti on we've implemented is to tip a rectangular steel block (approx. 1"x1"x3 1/2" ; longer would be better) on end and place it in the sash path next to the deflector shield.  If the sash is released, it will fall and be stoppe d by the block before crushing an arm.

 

Inspection can b e problematic depending upon where the counterweight is located.  There can be access and confined space issues.  Also, the failure occurred where the cable was stressed (crimped).  It may be more practical to replace cables regularly rather than attempt an inspection for defects that may not be visible.

 

Rick Clark

Sr. Research Chemist

Curwood, Inc.


From: DCHAS-L Discussion List [mailto:DCHAS-L**At_Symbol_Here**list.uvm.edu] On Behal f Of RJ Wolcik
Sent: Wednesday, November 11, 2009 12:26 PM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: [DCHAS-L] Hood Sash Cables

We are looking for some ide as on hood sash cables:

We inspect all our hood sas h cables quarterly by a visually inspecting the cable and running a finger al ong the cable to feel for any bumps or frayed wire.  We use 3/32” st ainless steel “aircraft grade” cables with a rating of 900 pounds.  ; We have found wire that has frayed in the middle of the twist and it is not visible witho ut a magnifying glass, and if it were to give away while someone was in the hood , we would have a serious injury.  We are thinking about instituting a chan ge schedule for all the cables. Does anyone have any ideas or comments?    

Does anyone have any inform ation on cable breakage, or has anyone heard of an incident of both sash cables breaking at the same time causing the sash to completely fall?

Does anyone have a policy t hey would be willing to share?

--------------------------- ------------------------

R.J. Wolcik

Environmental Health and Safety

Strem Chemicals, Inc.

7 Mulliken Way, Dexter Industrial Park

Newburyport, MA 01950-4098, U.S.A.

Telephone: 978.499.1660

Fax: 978.499.1661

Email: rj**At_Symbol_Here**strem.com

www.strem.com

--------------------------- -----------------------

Confident iality Notice: This e-mail transmission may contain confidential or legally privileged information that is intended onl y for the individual or entity named in the e-mail address. If you are not the intended recipient, you are hereby notified that any disclosure, copying, distribution, or reliance upon the contents of this e-mail is strictly prohibited. If you have received this e-mail transmission in error, please reply to the sender, so that Strem Chemicals Inc. can arrange for proper delivery , and then please delete the message from your Inbox. Thank you.


This email and a ny attachments may contain confidential and/or proprietary information. If you are not the intended recipient, you are not authorized to read, copy or use the contents of the email or any attachment. If you have received this email in error, please let us know by reply and then delete it from your system.

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.