ERPG-1, -2, and -3 values for it are: Hexafluoropropylene (116-15-4) 
10 ppm 50 ppm 500 ppm. Old-fashioned refrigerants look very good when 
human toxicity is factored in...

At 9:45 AM -0700 4/26/07, Russell Vernon wrote:
>Dear Fellow Chemists,
>
>
>
>In California we are being required to test some of our fume hoods with
>a tracer gas. We are also being encouraged to reduce our greenhouse gas
>emissions.  The standard method, ASHRAE 110-1995, calls our sulfur
>hexafluoride as the tracer gas of choice. It takes about 0.5 to 1 kg of
>tracer gas for each test. SF6 has 23,900 times the global warming
>potential of CO2 largely dues to its stability and very long residence
>times in the atmosphere. ASHRAE 110-1995 allows for alternative tracer
>gases to be used but requires that they are similar in weight, stability
>and detection limits to SF6.
>
>
>
>The best alternative I've found so far is hexafluoropropene [116-15-4].
>The current director of our Air Pollution Research Center
>(www.aprc.ucr.edu  ) assures me that
>hexafluoropropene will hang around in the atmosphere only 4-5 days so
>will not have the dramatic negative environmental impact of SF6 (also it
>will not deplete the stratospheric ozone).
>
>
>
>I have found a local fume hood certifier that is set up to conduct the
>ASHRAE 100-1995 test who is willing to try a new gas, but his detector
>is calibrated to SF6.
>
>
>
>We contacted the manufacturer of his Q200 leak detector and they need a
>certified calibration gas source. Normally they use a 1% mix of gas in
>nitrogen, and fill a 0.6 liter bottle with it to a pressure of between 4
>and 10 bar.  The set the leak rate of the gas at either 1 x 10-7 or 1 x
>10-6 cc/sec, which corresponds to 0.1 and 0.2 ppm respectively.
>
>
>
>Can you recommend a lab, supplier or manufacturer that would be willing
>to create a certified mixture of 1% hexafluoropropene in nitrogen?
>
>
>
>(The manufacturer of the Q200 is in the UK so a European source may be
>the best.)
>
>
>
>Sincerely,
>
>-Russ
>
>Russell Vernon, Ph.D.
>UC System-wide Field Safety Working Group Chair
>http://ehs55.ehs.uci.edu/fieldsafe/index.jsp
>
>Laboratory / Research Safety Specialist & Integrated Waste Manager
>Environmental Health & Safety
>University of California, Riverside
>900 University Ave.
>Riverside, CA 92521
>
>russell.vernon**At_Symbol_Here**ucr.edu
>www.ehs.ucr.edu
>
>Direct: (951) 827-5119
>Admin: (951) 827-5528
>Fax: (951) 827-5122
>
>

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.